CVE-2015-3956
Pifzer Plum A\+ Infusion System Firmware vulnerability
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
High priority: this vulnerability is rated critical and can be attacked over the network without a login.
Affected products
Plum A\+ Infusion System Firmware, Plum A\+3 Infusion System Firmware, Symbiq Infusion System Firmware. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Pifzer Plum A\+ Infusion System Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Limit access until thenIf you cannot update at once, take the system off the internet or restrict who can reach it.
References
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.