CONTENTUM.COMA specialist. A plan. A result.

Vulnerabilities · 2018

CVE-2018-14781
Medtronicdiabetes 508 Minimed Insulin Pump Firmware vulnerability

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Affected products

508 Minimed Insulin Pump Firmware, 522 Paradigm Real-time Firmware, 722 Paradigm Real-time Firmware, 523 Paradigm Revel Firmware, 723 Paradigm Revel Firmware, 523k Paradigm Revel Firmware, 723k Paradigm Revel Firmware, 551 Minimed 530g Firmware, 751 Minimed 530g Firmware. The exact versions are in the references.

What to do now

  • Check your exposureFind out whether Medtronicdiabetes 508 Minimed Insulin Pump Firmware runs anywhere in your environment, including systems of suppliers.
  • UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
  • Set the orderSystems that can be reached from the internet come first, then the rest.

References

Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.