CVE-2019-10964
Medtronic Minimed 508 Firmware vulnerability
Medtronic MiniMed Insulin Pumps are designed to communicate using a wireless RF with other devices, such as blood glucose meters, glucose sensor transmitters, and CareLink USB devices. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with adjacent access to one of the affected insulin pump models can inject, replay, modify, and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Affected products
Minimed 508 Firmware, Minimed Paradigm 511 Firmware, Minimed Paradigm 512 Firmware, Minimed Paradigm 712 Firmware, Minimed Paradigm 712e Firmware, Minimed Paradigm 515 Firmware, Minimed Paradigm 715 Firmware, Minimed Paradigm 522 Firmware, Minimed Paradigm 722 Firmware, Minimed Paradigm 522k Firmware and more. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Medtronic Minimed 508 Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Set the orderSystems that can be reached from the internet come first, then the rest.
References
- www.securityfocus.com/bid/108926 · Third Party Advisory
- www.us-cert.gov/ics/advisories/icsma-19-178-01 · Third Party Advisory
- global.medtronic.com/xg-en/product-security/security-bulletins/minimed-508-paradigm.html
- www.cisa.gov/news-events/ics-medical-advisories/icsma-19-178-01
- nvd.nist.gov/vuln/detail/CVE-2019-10964
- cve.org/CVERecord?id=CVE-2019-10964