CONTENTUM.COMA specialist. A plan. A result.

Vulnerabilities · 2019

CVE-2019-25719
Dräger Infinity Acute Care System and Standalone Infinity M540 …

Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.

What to do now

  • Check your exposureFind out whether the affected software runs anywhere in your environment, including systems of suppliers.
  • UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
  • Limit access until thenIf you cannot update at once, take the system off the internet or restrict who can reach it.

References

Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.