CVE-2019-25723
Dräger Perseus A500 software versions 2.00 through 2.02 contains an …
Dräger Perseus A500 software versions 2.00 through 2.02 contains an improper input handling vulnerability that allows external attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can overload the internal processor with malformed data to trigger a warm restart, causing ventilation pressure to drop to ambient level and interrupting ventilation for several seconds before therapy resumes.
What to do now
- Check your exposureFind out whether the affected software runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Limit access until thenIf you cannot update at once, take the system off the internet or restrict who can reach it.
References
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.