CVE-2020-10627
Insulet Omnipod Insulin Management System Firmware vulnerability
Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.
Affected products
Omnipod Insulin Management System Firmware. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Insulet Omnipod Insulin Management System Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Set the orderSystems that can be reached from the internet come first, then the rest.
References
- www.myomnipod.com/security-bulletins · Vendor Advisory
- us-cert.cisa.gov/ics/advisories/icsma-20-079-01 · Third Party Advisory
- nvd.nist.gov/vuln/detail/CVE-2020-10627
- cve.org/CVERecord?id=CVE-2020-10627
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.