Vulnerabilities · 2020 · Medtronic
CVE-2020-25183
Medtronic Mycarelink Smart Model 25000 Firmware vulnerability
Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.
Affected products
Mycarelink Smart Model 25000 Firmware. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Medtronic Mycarelink Smart Model 25000 Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Set the orderSystems that can be reached from the internet come first, then the rest.
References
More vulnerabilities in Medtronic products
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.