CONTENTUM.COMA specialist. A plan. A result.

Vulnerabilities · 2020 · Medtronic

CVE-2020-25187
Medtronic Mycarelink Smart Model 25000 Firmware vulnerability

Medtronic MyCareLink Smart 25000 is vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device

Affected products

Mycarelink Smart Model 25000 Firmware. The exact versions are in the references.

What to do now

  • Check your exposureFind out whether Medtronic Mycarelink Smart Model 25000 Firmware runs anywhere in your environment, including systems of suppliers.
  • UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
  • Set the orderSystems that can be reached from the internet come first, then the rest.

References

More vulnerabilities in Medtronic products

Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.