CVE-2025-53963
Thermofisher Ion Torrent Onetouch 2 Firmware vulnerability
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. They run an SSH server accessible over the default port 22. The root account has a weak default password of ionadmin, and a password change policy for the root account is not enforced. Thus, an attacker with network connectivity can achieve root code execution. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
High priority: this vulnerability is rated critical and can be attacked over the network without a login.
Affected products
Ion Torrent Onetouch 2 Firmware. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Thermofisher Ion Torrent Onetouch 2 Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Limit access until thenIf you cannot update at once, take the system off the internet or restrict who can reach it.
References
- documents.thermofisher.com/TFS-Assets/CORP/Product-Guides/Ion_OneTouch_2_and_Torrent_Suite · Vendor Advisory
- assets.thermofisher.com/TFS-Assets/LSG/manuals/MAN0014388_IonOneTouch2Sys_UG.pdf
- tools.thermofisher.cn/content/sfs/brochures/One_Touch_2_Spec_Sheet.pdf
- nvd.nist.gov/vuln/detail/CVE-2025-53963
- cve.org/CVERecord?id=CVE-2025-53963
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.