Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x …

Spacelabs Healthcare Sentinel versions 10.5.x and higher and 11.x.x before 11.6.0 contain an unauthenticated remote code execution vulnerability through a deprecated .NET Remoting HTTP channel exposed on port 8989 that allows attackers to perform arbitrary file read and write operations by supplying valid .NET URI endpoints. Attackers can write ASPX webshells to the IIS wwwroot directory to achieve unauthenticated remote code execution on the system. Port 8989 is not exposed in a default Sentinel installation; exploitation requires that the .NET Remoting port has been explicitly made network-accessible through deliberate configuration or network policy changes.

High priority: this vulnerability is rated critical and can be attacked over the network without a login.

What to do now

  • Check your exposureFind out whether the affected software runs anywhere in your environment, including systems of suppliers.
  • UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
  • Limit access until thenIf you cannot update at once, take the system off the internet or restrict who can reach it.