Vulnerabilities · 2014 · Baxter
CVE-2014-5431
Baxter Sigma Spectrum Infusion System Firmware vulnerability
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.
Affected products
Sigma Spectrum Infusion System Firmware. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Baxter Sigma Spectrum Infusion System Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Set the orderSystems that can be reached from the internet come first, then the rest.
References
More vulnerabilities in Baxter products
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.