Abbott Accent Firmware vulnerability
The authentication algorithm in Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017, which involves an authentication key and time stamp, can be compromised or bypassed, which may allow a nearby attacker to issue unauthorized commands to the pacemaker via RF communications. CVSS v3 base score: 7.5, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.
Affected products
Accent Firmware, Anthem Firmware, Accent Mri Firmware, Accent St Firmware, Assurity Firmware, Allure Firmware, Assurity Mri Firmware. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Abbott Accent Firmware runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Set the orderSystems that can be reached from the internet come first, then the rest.
References
- www.securityfocus.com/bid/100523 · Third Party Advisory
- ics-cert.us-cert.gov/advisories/ICSMA-17-241-01 · Third Party Advisory
- nvd.nist.gov/vuln/detail/CVE-2017-12712
- cve.org/CVERecord?id=CVE-2017-12712