CVE-2017-12725
Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump vulnerability

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

Affected products

Medfusion 4000 Wireless Syringe Infusion Pump. The exact versions are in the references.

What to do now

  • Check your exposureFind out whether Smiths-medical Medfusion 4000 Wireless Syringe Infusion Pump runs anywhere in your environment, including systems of suppliers.
  • UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
  • Limit access until thenIf you cannot update at once, take the system off the internet or restrict who can reach it.