Vulnerabilities · 2016 · Bd
CVE-2016-8375
Bd Alaris 8015 Pc Unit vulnerability
An issue was discovered in Becton, Dickinson and Company (BD) Alaris 8015 Point of Care (PC) unit, Version 9.5 and prior versions, and Version 9.7, and 8000 PC unit. An unauthorized user with physical access to an affected Alaris PC unit may be able to obtain unencrypted wireless network authentication credentials and other sensitive technical data by disassembling the PC unit and accessing the device's flash memory. The Alaris 8015 PC unit, Version 9.7, and the 8000 PC unit store wireless network authentication credentials and other sensitive technical data on internal flash memory. Accessing the internal flash memory of the affected device would require special tools to extract data and carrying out this attack at a healthcare facility would increase the likelihood of detection.
Affected products
Alaris 8015 Pc Unit. The exact versions are in the references.
What to do now
- Check your exposureFind out whether Bd Alaris 8015 Pc Unit runs anywhere in your environment, including systems of suppliers.
- UpdateInstall the version in which the vendor fixed the vulnerability. The references lead to the advisory.
- Set the orderSystems that can be reached from the internet come first, then the rest.
References
More vulnerabilities in Bd products
Data: CVE list and the NVD. This product uses data from the NVD API but is not endorsed or certified by the NVD.